THE BETTER MOUSE TRAP CO. · COMPLETED METHOD RECORD

Risk matrix and SWOT — D1 pilot decision

DocumentBMT-RISK-MATRIX-AND-SWOT-D1
Revision / review dateD1 · 2027-06-30
Accountable ownerAlex Meadow
D1 residual severity/occurrence grid; severe flight exposure remains unacceptable despite bench results.9–107–85–63–41–21–23–45–67–89–10H01, D01B01D02: flight deniedSeverity ↑ · Occurrence → · residual D1 ratings; colors require review, never permission
D1 residual severity/occurrence grid; severe flight exposure remains unacceptable despite bench results.

Scope: Risk matrix and SWOT — D1 pilot decision. Read the observations and decisions below within their stated conditions.

Risk register · 1–10 severity and occurrence

ID / hazardInitial S/OD1 residual S/OControl / decision
B01 rebound7/57/2Key hinge, 0/1000; accept controlled pilot
H01 missed local alert9/59/2Retained state, 0/1000; supervised bench
D01 unsafe enable10/410/2Hardware inhibit, 30/30; bench only
D02 flight exposure10/610/6Not reduced by bench tests; flight denied

SWOT · evidence, response and owner

QuadrantEvidence / response / owner
Strength 1Serial-to-test trace for all three products; Ira retains independent evidence review.
Strength 2Basic keyed hinge achieved 0/1000 rebounds; Sam retains functional retention checks.
Strength 3Hotel local state works offline; Jordan retains power-loss regression.
Strength 4Drone motor isolation separates HIL from flight; Ira witnesses before every run.
Weakness 1Surrogates cannot establish animal welfare; Robin denies live-use claims.
Weakness 2Small prototype sample cannot establish production capability; Sam excludes capability claims.
Weakness 3No approved flight hardware or site; Alex denies flight and field release.
Weakness 4No external RF or battery qualification; Jordan keeps cabled/dummy-load configuration.
Opportunity 1Station gauge can reduce repeat setup; Sam issued fixture-readiness WI June 30.
Opportunity 2Offline evidence index prevents stale instruction use; Alex verified 4/4 stations.
Opportunity 3Independent reviewer can catch missed CTQ traces; Ira checked D1 chain.
Opportunity 4User task protocol can improve manual recovery; Robin reviewed 10/10 surrogate recoveries.
Threat 1Bench success could be marketed as safety approval; Alex denied consumer claims.
Threat 2Firmware bypass could defeat inhibit; Jordan enforced signed-image/profile check.
Threat 3Power loss could hide occupancy; Jordan retained local flag and manual access.
Threat 4Unqualified field release could expose animals and people; Robin/Alex denied destination and flight approval.

Approval

AuthorityOutcome
Alex signed June 25; Ira checked plotting and evidence June 20Accept only listed internal scope. Color and RPN never override severe-hazard review.

Record reviewed and signed: Alex Meadow, 2027-06-30. Technical review: Sam Reed / Jordan Vale; independent check: Ira Bell. Disposition: use for the stated internal scope only.

Compare the reference example. Teaching data and signed dispositions: D1 scenario register.